LYSAM-SN Stationary Automatic Water Sampler

Stationary Automatic Sampler
LYSAM-SN is an outdoor stationary automatic water sampler that uses a vacuum water sampling method. It is operated using a color touch panel, and sampled water is stored in a refrigerator. In addition, water sampling history and alarm history can be retrieved using a USB memory stick.Optional features include auto-drain water sampling and high-head water sampling up to 30m.*The control unit window is optional.
Main function
1. Timing of water collection and method of distributing the sample into bottles Automatic water sampling by programming
2. Setting the amount of water to be collected at one time
3. Starting the water sampling program by external input signal
4. Controlling the progress of the water sampling program using external input signals
5. Contact output for alarm status, operation status, etc. of this device (JS specification)
6. Save 5 types of water sampling programs
7. Water sampling history and alarm history record
8. SDI-12 communication (optional)
9. Refrigerate and store collected water samples
specification
Water sampling target
General sewage, rainwater, industrial wastewater, sludge, rivers, lakes, etc.
Main unit
| Main material | SUS304 (standard), SUS316 (optional) |
| size | W800×H1750×D750 (excluding protrusions) |
| display | TFT color touch panel |
| Ambient temperature | 0 to 50°C |
| Ambient humidity | 25 to 95% RH (no condensation) |
function
| Water sampling method | Diaphragm pump type (vacuum water sampling) |
| Water collection speed | 0.9m/s or more with horizontal pull of 3m and lifting height of 6.2m |
| Water sampling accuracy (reproducibility) | Within ±2% |
| Water intake head | V1: 8.5m V2: 9.5m (single suction) V3: 20m V4: 30m (double suction) V5: 18m (hybrid) |
| Measuring part | 500mL chamber and variable volume top (1L and 2L are optional) |
| Sample bottle | 2L x 24 bottles, 30L x 1 bottle |
| Water sampling hose(sampling line) | Soft PVC black braided hose ID9.5 x OD16 connector with strainer |
| temperature display | Thermistor type water temperature sensor (optional) |
| Data logger | Records water sampling time, mode, bottle number, temperature inside the chamber, water sampling results, analog instantaneous values, etc., and collects data via USB memory |
| input | Touch panel programming |
| timer | Delay or real-time timer |
| Water sampling mode | Time proportional water sampling, flow proportional water sampling (analog), irregular interval water sampling |
| Water sampling control | External water sampling control (analog, contact, water temperature (optional)) Water sampling control by day of the week Irregular interval water sampling |
| Water sampling interval | Time proportional: 1 to 9999 (unit: min) Flow proportional: 1 to 99999 (unit: m3) |
| Number of divided water samples | 2 to 999 times |
| Distribution method | Composite water sampling, single water sampling, multi-sample water sampling, multi-bottle water sampling, day-step water sampling, auto-drain water sampling (optional) |
| Cycle Settings | Water collection cycle and retry count settings |
| Temperature recording function | Logging of water temperature at each sampling point (optional) |
Analog Signal Input
DC4 to 20mA (standard 1 point, maximum 2 points)
flow rate, water quality, etc.
Abnormality signal input
No-voltage contact signal
power supply
AC100V 50/60Hz
Power consumption
Standard 550W, high head 800W
(rush current 50Hz 23A, 60Hz 22A)
Body mass
Approximately 300kg
Standard accessories
| Water collection bottle | 2L x 24 bottles (stored inside the bottle case) |
| Water sampling hose | 10m (with connector and strainer) 1 set |
| Composite Bottle | 30L x 1 bottle |
| rubber seal | 1 pair |
option
- Sample bottles (2L x 24 bottles, 4L x 12 bottles, 8L x 4 bottles, 10L x 2 bottles, 30L x 1 bottle, etc.)
- Chamber and top (1L, 2L)
- Water sampling start level detector
- SDI-12 communication capability
- High lift specifications (20m/30m)
- Anti-siphon function
- Polyurethane coating
- AC200V specification
- Strainer (SUS304)
- Water Temperature Sensor
- Auto-drain water collection function
- Intermediate tank lifting device
- Hose cleaning function
- Salt damage prevention specifications
- Casters (with stoppers)
- Control room window
- Analog signal (2 points)
Installation conditions
Since the LYSAM series uses a vacuum water sampling method, the sampling point must be installed at a lower position than the chamber. Install the water sampling hose so that it does not slope backward from the chamber to the sampling point.
Disposal
Please contact us for a fee. *Refrigerant recovery is also available.
Recycled Design
The LYSAM series aims to create environmentally friendly products, and is mindful of reducing, reusing, and recycling.*Please note that specifications may be changed without notice for improvement purposes.*The shape and performance of the product may change depending on the installation environment, construction conditions, etc.